Backreflection.

Backreflection monitors light attenuation due to scattering to assess aggregation and measure sample turbidity.

When to use backreflection.

The technique measures a sample’s turbidity, or cloudiness. As proteins aggregate, they scatter light away from the detection path, reducing the backreflected signal. This provides a sensitive readout of aggregate formation during thermal or chemical stress.

Backreflection is especially useful for proteins that form large, visible aggregates during thermal unfolding. It’s less sensitive to small oligomers or subtle changes in particle size, where DLS excels.

Backreflection is ideal when you need to:

Solution

Detect aggregation during thermal stress.

Solution

Identify conditions that minimize aggregation.

Solution

Distinguish between unfolding and aggregation (when combined with nanoDSF).

Solution

Monitor large aggregate formation that might not be visible by DLS.

Solution

Screen formulations for colloidal stability.

The science behind backreflection.

The science behind DLS Image

Aggregates scatter light and reduce backreflected signal

In a backreflection measurement, light is directed into the sample and reflects off a mirror positioned behind the sample volume. The reflected light passes back through the sample and returns to the detector.

When the sample contains well-folded, monomeric protein, most of the light passes through with minimal scattering. The backreflected signal remains high.

As aggregates form—whether from thermal stress, chemical denaturants, or prolonged incubation—they grow large compared to the folded protein. Large aggregates scatter light at wide angles, deflecting it away from the detection path. This reduces the intensity of light reaching the detector.

The decrease in backreflected light intensity is a direct measure of turbidity—the cloudiness of the solution.

Turbidity increases with aggregate size and concentration

Turbidity depends on both the size and number of aggregates. Small aggregates scatter less light than large ones. As aggregation progresses, both the size and number of aggregates increase, causing turbidity to rise.

Backreflection is particularly sensitive to large, amorphous aggregates that scatter light efficiently. This makes it complementary to DLS, which is better suited for detecting smaller aggregates and measuring their size distribution.

Parameters you measure with backreflection.

Scientist operating the Prometheus Panta instrument in a lab

Onset of aggregation (Tagg or Tturb)

The temperature at which the backreflected signal begins to decrease, indicating the onset of aggregation. This parameter identifies when the sample starts to lose colloidal stability.

Aggregation profile

The shape and magnitude of the turbidity signal over the course of the experiment. This reveals whether aggregation is rapid or gradual, and whether it correlates with protein unfolding.